Chemistry, 21.01.2021 21:50 jatavispillow
The detection limit for caffeine analysis must be calculated. Eight absorbance readings from a caffeine sample close to the estimated detection limit are 0.1239, 0.1098, 0.1182, 0.1150, 0.1219, 0.1251, 0.1199, and 0.1099. The average absorbance for eight blanks is 0.0182. The slope of a calibration curve is 0.59 /mM. What is the detection limit
Answers: 2